    Accuracy

    =1%@100% stroke

    Repeat accuracy 

    3‰ - 7‰

    Pressure

    0-1.0Mpa (air)

    0-1.2Mpa (water)

    Service life      

    3 million times no leakage (media: water)

    Detection

    Photosensor check original piston position for accurate location

    Rated stroke (control steps) 

    5ml

    10ml

    20ml

    30mm  (12000 steps)

    24.08mm  (9632 steps)

    24.88  (9952 steps)

    Linear speed

    0.017mm/s – 5mm/s

    Time range (rated stroke)

    6s – 1765s

    4.8s – 1416s

    5s – 1464s

    Accuracy resolution 

    0.0025mm/0.4154µL

    0.0025mm/1.0381µL

    0.0025mm/2.0096µL

    Actuator

    Ball screw (lead 1mm)

    Max piston drive     

    = 100N

    Sub piston drive    

    = 45 N

    Syringe ID

    5ml

    10ml

    20ml

    14.55mm

    23mm

    32mm

    Fluid path         

    Borosilicate cylinder, PTFE piston

    Max pressure          

    Positive air pressure 0-1.0Mpa / Negative air pressure 0-0.05Mpa

    Connector

    1/4-28UNF female

    Communication

    RS232/RS485/CAN

    Baud rate            

    9600dps, 19200dps, 38400dps, 57600dps, 115200dps

    Device add. & Parameter setting 

    Serial Pont

    Dimension (L*W*H)

    42*42*210mm

    Power supply      

    DC24V/1.5A

    Working temperature           

    5 – 55?

    Working humidity

    < 80%

    Weight 

    0.72kg


    Product Function

    Address setup

    Serial port

    Baud rate setup

    Different baud rates under RS232/RS485/CAN available

    CAN target add setup

    Superior device setup support

    Speed setup

    Serial setup 1-250rpm/min (air and liquid different)

    Subdivision setup

    Serial setup: subdivision must be 256 when speed as 1rpm/min

    Reset internal data

    Factory setting

    Inquiry parameter

    Device add, speed, subdivision, baud rate

    Inquiry version

    Current version

    Motor direction

    CW, CCW

    Reset

    Return piston to origin

    Strong stop

    Strong stop the running motor

    Motor status inquiry

    Inspect current motor status

    Power memory

    When motor suddenly stops, current position can be enquiry by display the distance to origin


    Motor Parameter

    Name

    0.9°step motor

    1.8°step motor

    Max power

    9.2W

    9.2W

    Step angle

    0.9°

    1.8°

    Phase

    2

    2

    Phase volt

    4.6V

    4.2V

    Phase current

    1.0A

    1.1A

    Resistance

    4.6O±0.48

    3.8O±0.38

    Inductance

    18.6mH REF

    5.2mH REF

    Insulation

    100m O MIN

    100m O MIN

    Max Temp.

    80? MAX

    80? MAX

    Explosion-Proof level

    B

    B


    Motor Wireline

    Definition

    0.9°step motor line

    1.8°step motor line

    A

    Orange

    Orange

    A-

    Blue

    Blue

    B

    Red

    Red

    B-

    yellow

    yellow


    Driver Port Direction

    image003.jpg

    Port Name

    Description

    Port Name

    Description

    +

    DC24V positive

    A-

    step motor A-

    -

    DC24V negative

    B+

    step motor B+

    TX

    RS232 TX

    B-

    step motor B-

    RX

    RS232 RX

    O1

    IO1

    GND

    RS232 Land

    O2

    IO2

    H

    CAN - H

    O3

    IO3

    L

    CAN - L

    V1

    V1

    A

    RS485 - A

    V2

    V2

    B

    RS485 - B

    V3

    V3

    A+

    step motor A+

    GND

    earth wire


    Connection Way

    image005.jpg

    Working Flow

    image007.jpg


Send your message to this supplier
To:  

Nanjing Runze Fluid Control Equipment Co.,Ltd

* Content:  

For better quotations, include:

- A self introduction

- Special requests, if any

Quantity:  

  
You May Like
Not exactly what you want? One request, multiple quotations Get Quotations Now >>
Related Searches :